The Fluidex oil moisture meter features a compact design and ease of operation — changing the sample requires only replacing the removable container and pressing the Start button to begin measurement. The device is multifunctional and does not require a personal computer for operation. Its protective design, including a stabilized power supply, metal vandal-resistant buttons, safeguarded sensing elements, and integrated protection of all electrical components, allows it to be used on site near the transformer. Measurement results can be immediately printed out using an integrated printer.
Categories : Oil Testing
FLD 1T Oil Moisture Meter
The FLD 1T transformer oil meter is engineered to determine the mass fraction of dissolved moisture in insulating mineral oils with a kinematic viscosity of up to 70 cSt at 50 °C.
When using the unit, it is important to keep in mind that the FLD 1T provides a rapid indicative measurement of moisture content. Obtaining definitive, certified results still requires a comprehensive laboratory analysis.
-
Description
The FLD 1T transformer oil meter is engineered to determine the mass fraction of dissolved moisture in insulating mineral oils with a kinematic viscosity of up to 70 cSt at 50 °C.
When using the unit, it is important to keep in mind that the FLD 1T provides a rapid indicative measurement of moisture content. Obtaining definitive, certified results still requires a comprehensive laboratory analysis.
Conventional laboratory testing typically takes at least one day and requires considerable labor and financial resources. The FLD 1T enables immediate assessment of transformer oil condition directly on site, allowing operators to make prompt, informed decisions near the transformer without delays.
The device integrates all the components required for easy and user-friendly operation, including:
- LCD display that shows the measured moisture content;
- intake container for collecting an oil sample;
- moisture sensor that determines the mass fraction of water in lubricating and insulating oils;
- integrated printer for generating hard-copy test reports.
-
Specification
No. Parameter
Value
1 Operating range for active water, aw 0…1
2 Operating range for dissolved water content measurement, ppm 0….500
3 Precision at 68 °F (20 °C) ±0.02 aw (0…0.9 aw) ±0.03 aw (0.9…1 aw)
4 Temperature measurement yes
5 Operating medium temperature, °F (°С) 32…131 (0…55)
6 Required power, W 10
7 Single-phase 60Hz (50Hz) AC power supply voltage, V 120 (230)
8 Dimensions, in (mm) – length
– width
– height
8 (190)
10 (255)
14 (365)
9 Weight, lbs (kg) 13.2 (6)
-